ECN-2 motion film processing with the ECN-PRO3 processor

ECN-2 Process Control: How to Read Control Strips and Plots

Short answer: an ECN-2 lab keeps its process in control by running pre-exposed control strips through the processor at regular intervals, reading them on a densitometer and plotting the results against aim values. Kodak's process control manual (publication H-24.01) recommends, as a rule of thumb, one strip every hour of operation.

Chemistry ages, replenishment drifts and temperatures wander. Without control strips, a lab only finds out when a customer's negative comes back thin or off-colour. With them, the change shows up on a plot before it reaches anyone's footage, whatever processor the lab runs.

What a control strip is

Kodak supplies sensitometrically exposed control strips for Process ECN-2. Kodak's process control module describes a package of one 30.5 m (100 ft) roll with at least 120 exposures, together with a reference strip that Kodak has processed. Each strip carries 21 grey-scale steps at 0.20 log H increments: 20 exposed steps plus a minimum-density area.

Kodak advises that where a lab runs more than one processor for the same process, each machine should have its own control strips, so that a problem on one machine cannot be hidden by another.

The patches that matter in ECN-2

Kodak's manual names the readings used for ECN-2, all taken with Status M densitometry:

Reading ECN-2 patch What it tells you
D-min Opposite dot Minimum density
LD (low density) Step 14 Toe
MD (mid density) Step 8 Speed
HD (high density) Step 4 Picture blacks
HD – LD Steps 4 to 14 Contrast

Colour balance is read as the spread between the red, green and blue readings within a step. Kodak notes that the HD patch represents the picture blacks but not necessarily maximum density, so ECN-2 control does not use a D-max reading.

Aim values, action limits and control limits

When a new batch of strips is opened, Kodak's method is to read the reference strip and apply the correction factors printed on that batch's instruction sheet. The results are the aim values, which become the zero line on the control plot.

Kodak's recommended tolerance limits for ECN-2 are:

Reading Action limit Control limit
D-min +0.03 +0.05
LD (step 14) ±0.03 ±0.05
MD (step 8) ±0.05 ±0.07
HD (step 4) ±0.06 ±0.08
HD – LD ±0.05 ±0.07

At the action limit, take corrective action. At the control limit, Kodak advises suspending processing until the precise cause has been found. Kodak describes the limits themselves as arbitrary, so a lab may set its own, but once agreed they should not be changed without the agreement of everyone who relies on them.

Running, storing and reading strips correctly

A control plot is only as reliable as the strips and the densitometer behind it. Kodak's guidance includes:

  • Storage. Keep unused strips at −18°C (0°F) or lower. Take out one day's supply in the dark, keep it in a light-tight container below 27°C (80°F), and discard what is left at the end of the day.
  • Warm-up. Let a can reach room temperature before opening it, which may take half an hour to an hour, and never use heat.
  • Handling. Handle strips only in total darkness and by the edges, and never use them beyond their expiry date.
  • Densitometer. Calibrate at the start of each working day, check against a standard plaque, read the centre of each patch, and read control and reference strips on the same instrument.

Kodak's 2025 guide to using kit chemicals in motion picture laboratories adds that its replenishment rates are starting points, to be adjusted according to control plot results.

Reading the plot: common signals

Trends tell you more than single points. Kodak's guidance gives examples a lab will recognise:

  • Rising red and green D-min. Kodak's kit chemicals guide links an increase in red and green D-min on the control plot to incomplete fixing. Re-fixing the strip in known-good fixer is a quick test: if the plot corrects, the original fixer is probably exhausted.
  • A shift after a shutdown. Kodak notes that a trend after a long shutdown can be a start-up effect that running additional leader may stabilise.
  • A change with no obvious process cause. Kodak warns that changes can also appear if strips are poorly stored or solutions are mixed with inferior chemicals, so check the strips and the mix before adjusting the process.

Micrographics Data makes the ECN-PRO for remjet stocks and the ECN-PRO3 for remjet-free (AHU) stocks. Whichever machine a lab runs, the same control-strip discipline keeps its results consistent.

Frequently asked questions

How often should an ECN-2 lab run control strips?

Kodak's process control manual suggests, as a rule of thumb, processing one strip every hour, and notes that experience will determine the frequency a lab ultimately needs.

Which densitometer filters are used for ECN-2 control strips?

Kodak specifies Status M densitometry for ECN-2 control strips.

What should a lab do when a reading passes the control limit?

Kodak advises suspending film processing until the precise cause has been found. At the action limit, the lab should take corrective action.

How should unused control strips be stored?

Kodak advises storing unused strips at -18 degrees C (0 degrees F) or lower, and keeping each day's supply in a light-tight container below 27 degrees C (80 degrees F).

Do the ECN-PRO and ECN-PRO3 need different process control?

No. Both run Process ECN-2, so the same control strips, aim values and limits apply. The ECN-PRO is designed for remjet stocks and the ECN-PRO3 for remjet-free (AHU) stocks.